Dew point set FA 410 for membrane and adsorption dryers
|
| | Special features • System ready for plug-in: everything completely wired • No time-consuming studying of the instruction manual • 2 alarm contacts (230 VAC, 3 A) • pre- and main-alarm freely adjustable • Optical alarm: Red flashing • Parallel indication % RH, °C, °Ctd • USB- Schnittstelle zum PC • Data logger for 1 mio. measuring values (optional)
Data sheet dew point set FA 410 pdf (82KB)
|
| | DS 300
| | | | Dimensions wall mounting: | 118 x 115 x 93 mm | | | Dimensions panel mounting: | 92 x 92 mm | | | Inputs: | 2 digital inputs for FA 410 resp. VA 400 | | | Power supply: | 100 ...240 VAC, 50-60 Hz | | | Interface: | USB | | | Operating temperature: | 0 ...50°C | | | Alarm output: | 2 relays (230 VAC, 3A) | | | Length power cable: | 3 m | | | Dew point sensor FA 410
| | | | Measuring range: | -80...20 °Ctd resp. 0...100 %RH | | | Pressure range: | -1...50 bar | | | Operating temperature: | -20...70 °C | | | Connection: | M12, 5-pole | | | Screw-in thread: | G1/2” | | | Dimensions: | Ø30 mm, length 130 mm | | | Length sensor cable: | 5m |
| | DESCRIPTION
| ORDER-NO.
| | | Dew point set FA 410 for adsorption dryers (-80...20 °Ctd) consisting of: | 0600 0410 | | | Multifunction measuring instrument DS 300 | 0500 3000 | | | FA 410 dew point sensor -80 ...20°Ctd | 0699 0410 | | | Standard measuring chamber up to 16 bar | 0699 3390 | | | Options:
| | | | Integrated data logger for 1 million measured values | Z500 3001 | | | Alarm unit mounted at wall housing | Z500 0003 | | | Alarm unit for external mounting with 5 m cable | Z500 0004 | | | Special version FA 410 up to 350 bar | 0699 4003 | | | Output in mg/m3, g/kg, oder %RH, special scaling 4...20 mA | 0699 4004 |
| | Description Accessories
| ORDER-NO. | | | Connection cable FA/VA series 400, 5m with M12 plug, open ends | 0553 0104 | | | Connection cable FA/VA series 400, 5m with M12 plug, open ends | 0553 0105 | | | Alarm cable, length: 5m | 0553 0106 | | | Alarm cable, length: 10m | 0553 0107 | | | High-pressure measuring chamber up to 350 bar (in conjunction with FA 410 for 350 bar) | 0699 3590 | | | Measuring chamber for atmospheric dew point | 0699 3690 | | | Measuring chamber for respiratory air bottles up to 350 bar | 0699 3790 | | | Measuring chamber for granulate dryers | 0699 3490 | | | CS Soft Professional for data evaluation incl. USB interface | 0554 7010 | | | CS Service Software VA/FA incl. PC connection set and CS Soft Professional | 0554 2005 | | | Precision calibration at -40°Ctd including ISO certificate | 0699 3396 | | | Precision calibration at 0 °C and 10 °Ctd including ISO certificate | 3200 0003 | | | Control and calibration set 11.3 % RH | 0554 0002 | | | Control and calibration set 33 % RH | 0554 0004 | | | Control and calibration set 75.3 % RH | 0554 0005 |
 | |  | |  | |  | |  | | Standard measuring chamber
| | High-pressure measuring chamber | | Measuring chamber for atmospheric dew point | | Measuring chamber for respiratory air bottles | | Measuring chamber for granulate dryers | |  | | | | | | | | | | Control and calibration set
| | | | | | | | | |
|