Dew point sensor FA 400 replacement for FA 200
|
| | Special features • Measuring range -60 ...+30 °Ctd • Integrated display • Threshold value adjustable via keypad alarm relay (max. 60 VDC, 0.5 A) • Extreme long-term stability • Quick response time • 4...20 mA analogue output • Pressure tigth up to 350 bar (special version) FA 400 features a complete dew point meter with sensor, display, keybord and alarm. The keyboard is placed behind a rubber cover and can be accessed by opening the cover on the top. Via two keys alarms and display features can be set . Sensor parameters such as analogue output scaling, alarm values, units etc. can be easily changed by using the CS Service software. This Software combined with the connection set is used to connect the sensor to a PC for configuration changes Recommendation: Mounting with standard measurement chamber for compressed air up to 16 bar Advantage: Easy installation via fast coupling .
Data sheet FA 400 pdf ( 58 KB) |
| | Measuring range: | -60...30 °Ctd resp. 0...100 %RH | | | Accuracy:
| ± 1 °C at 20...-20 °Ctd ± 2 °C at -20...-50 °Ctd ± 3 °C at -50...-60 °Ctd | | | Pressure range: | -1...50 bar, special version up to 350 bar | | | Stromversorgung: | 24 VDC (16...30 VDC) smoothed | | | Protection class: | IP 65 | | | EMC: | according to DIN EN 61326 | | | Operation temperature: | -20...70 °C | | | Connection: | 2 x M12, 5-pole for analogue output and alarm output | | | PC connection: | SDI interface | | | Output: | 4...20 mA = -60...+30 °Ctd, 3-wire technology | | | Burden for analog output: | < 500 Ohm | | | Alarm output: | NO, max. 60 VDC, 0.5 A | | | Screw-in thread: | G1/2” | | | Dimensions : | Ø65 mm, length approx.. 160 mm | | | Output signals via software: | %RH, °Ctd, g/m³, mg/m³, ppm V/V |
| | DESCRIPTION
| ORDER-NO.
| | | FA 400 dew point sensor, replacement of FA 200, -60 ...30°Ctd | 0699 0403 | | | Connection cable for FA/VA series 400, 5m with M12 plug, open ends | 0553 0104 | | | Connection cable for FA/VA series 400, 10m with M12 plug, open ends | 0553 0105 | | | Alarm cable, length: 5m | 0553 0106 | | | Alarm cable, length: 10m | 0553 0107 | | | Cable for FA 200 to FA 400 for feed-in and analogue output, length 1m | 0553 0112 | | | Cable for FA 200 to FA 400 for alarm output, length 1m | 0553 0113 | | | Standard measuring chamber for compressed air up to 16 bar | 0699 3390 | | | Options: | | | | Special version FA 400 up to 350 bar | 0699 4003 | | | Special scaling, output in ppm V/V, % RH, mg/m3 | 0699 4004 |
| | DESCRIPTION ACCESSORIES
| ORDER-NO. | | | High-pressure measuring chamber up to 350 bar (in conjunction with FA 410 for 350 bar) | 0699 3590 | | | Measuring chamber for atmospheric dew point | 0699 3690 | | | Measuring chamber for respiratory air bottles up to 350 bar (in conjunction with FA 410 for 350 bar) | 0699 379 | | | Measuring chamber for granulate dryers | 0699 3490 | | | CS Service Software for FA/VA 400 sensors including PC connection set, USB adapter and interface adapter to the sensor as well as CS Soft Professional software for data recording | 0554 2005
| | | Precision calibration at -40°Ctd including ISO certificate | 0699 3396 | | | Precision calibration at 0 °C and 10 °Ctd including ISO certificate | 3200 0003 | | | Mains unit in wall housing fpr sensors of the Series VA/ FA, 100.240 V, 10 VA, 50-60 Hz/ 24 VDC, 0,35 A | 0554 0108 | | | Control and calibration set 11.3 % RH | 0554 0002 | | | Control and calibration set 33 % RH | 0554 0004 | | | Control and calibration set 75.3 % RH | 0554 0005 |
 | |  | |  | |  | |  | | | Standard measuring chamber | | High-pressure measuring chamber | | Measuring chamber for atmospheric dew point | | Measuring chamber for respiratory air bottles | | Measuring chamber for granulate dryers | |
|