Details
Description of the measuring instrument:
The OIL CHECK 511 Point-of-Use Monitor is an at-line sensor. It continuously draws a small sample stream and measures the residual oil content present in the vapour-phase.
To reliably monitor critical processes and plant components, measurement after media treatment is often insufficient. Situations can also arise within the distribution network that cause oil components to enter the system. The OIL CHECK 511 helps detect such situations through changes in the measured concentrations of gaseous oil components. In this way, it helps protect critical processes.
- Indicative measurement according to ISO 8573-5:2025: Indicative measurement according to ISO 8573-5 provides transparency exactly where it matters most in operation: directly at the point of use or centrally in the compressed air system—and without having to wait for laboratory results. It continuously monitors residual oil levels and immediately highlights changes in the process, ensuring greater safety and faster response times. It focuses on gaseous components—particularly oil vapours—and thus complements quantitative laboratory analysis with practical online monitoring. This provides operators with a powerful tool for consistent compressed air quality, reliable processes, and operational safety.
- 7 calibration points, 1 of which is within Class 1: The most important factor for monitoring the purity grade is the correct calibration of the measurement system. The Class 1 limit value specified in ISO 8573 is 0.01 mg/m³. This corresponds to a volume concentration of 2.5 ppb. That means 2.5 particles per billion. The OIL CHECK 511 is calibrated with one adjustment point in Class 1, precisely where it is critical for the transition to Class 2. Calibration at 6 additional points ensures that even high residual oil levels can be reliably measured.
- Very easy installation and maintenance, even under pressure: The OIL CHECK 511 is designed for quick replacement in the field. Thanks to its compact design, low weight, quick coupling, and M12 connector, the sensor can be easily installed and removed. This allows it to be quickly replaced with a spare sensor for service or recalibration. As a result, measurement remains available with minimal effort.

- Low cross-sensitivity to humidity and ethanol: Cross-sensitivity to gases that are not target gases for oil vapour measurement can increase the measured value even when no relevant contamination is present in the measurement gas. The OIL CHECK 511 is specifically designed to have low cross-sensitivity to the two commonly present interfering gases, humidity and ethanol. This improves the reliability of the measured values.
Frequently Asked Questions (FAQ) About Residual Oil Measurement with the OIL CHECK 511
From a customer’s facility: In a semiconductor manufacturing plant, a purity grade of 1 according to ISO 8573-1 had been defined for the compressed air supply. It was not until the quality of the manufactured products no longer met the specified process parameters that troubleshooting began. The cause was not the central compressed air generation system, but rather an air preparation unit at the point of use: An oiler had been filled there, even though oiled compressed air was not approved at that location. As a result, the required compressed air quality was no longer maintained at the point of use.
This is precisely where continuous point-of-use measurement of compressed air quality proves its value: It reveals deviations where they occur, documents the status during ongoing operation, and helps pinpoint the cause more quickly before rejects, production downtime, and other associated costs increase. For sensitive processes, such as those in semiconductor manufacturing, continuous residual oil measurement at the point of use is therefore a crucial component for reliably ensuring that specified process parameters are met.
From a customer’s facility: During the installation of a new compressed air supply system, the OIL CHECK 500 was used to test the media transfer from the temporary supply station into the compressed air network. Even before the system was connected to the main network, the device indicated an extremely high measured value. The cause lay with the temporary station: Pipe connections had been made using an adhesive not approved for purity grade 1. Even flushing with clean air was not sufficient to remove the contamination.
This is precisely where continuous point-of-use measurement at media transfer points proves its value: It reveals deviations in compressed air quality exactly where quality must be right—right where it is delivered. This protects the downstream compressed air network from the ingress of residual oil or other contaminants before production equipment or process parameters are compromised.
This is also the case in industrial parks. There, a point-of-use monitor is used to test and document media quality in order to safeguard in-house production.
The OIL CHECK 511 is designed as an at-line monitor.
- Design: Thanks to its compact design and low weight, it can be easily integrated into existing systems. It can also be easily incorporated into the design of new machines.
- Data Transmission: With a Modbus RTU connection and a 4 … 20 mA connection, the OIL CHECK 511 features the essential interfaces for field devices.