VC 600 Clamp-on Flow Sensor

Measuring compressed air and gases without opening the pipe

The VC 600 is a clamp-on flow meter for the consumption measurement of compressed air, nitrogen and other gases. It uses ultrasonic technology to measure from the outside and is clamped onto the pipe using a quick clamp system; the pipe remains closed and under pressure. This makes it suitable for temporary measurements during audits and, in particular, for points of measurement where drilling or opening the pipe is not required.

Advantages of the VC 600:

  • Installation without interfering with the pipe: The quick clamp system, consisting of two half-shells with four quick-release clamps, is attached to the pipe – without drilling, welding or a ball valve; the system continues to operate. No measuring element protrudes into the cross-section, so there is no additional pressure loss.
  • No ultrasonic gel, no calibration work: Simply place the sensor unit onto the quick clamp system, secure it with two screws, and take the measurement.
  • One device for audits and continuous measurement: The same sensor unit can be used alternately at various points of measurement on different quick-clamping systems, or it can remain on the pipe where continuous measurement is required.
  • One sensor unit for all pipe diameters from 25 to 90 mm: The quick clamp system is selected to suit the existing pipe size; the sensor unit remains the same. Larger diameters available on request.
  • For steel, stainless steel and aluminium with wall thicknesses of 1.5 to 5.0 mm, including thin-walled stainless steel pipes in accordance with DIN EN 10357 and aluminium pipe systems in compressed air networks.
  • Standardized volume flow displayed directly on the unit: enter the inner diameter, wall thickness, pressure and temperature, and the sensor automatically calculates the standardized volume flow in Nm³/h in accordance with ISO 1217 or DIN 1343. No separate flow computer is required; the meter reading and chart recorder are integrated.
  • From 3 bar(g) line pressure, with no upper pressure limit imposed by the sensor: as the pressure rises, the measurement signal becomes stronger; this makes the VC 600 suitable for high-pressure networks as well.
  • Choice of interfaces: 4…20 mA in every version, plus Modbus RTU, IO-Link, Modbus TCP with Power over Ethernet or M-Bus. Operation via the touchscreen or via Bluetooth using the CS SensorApp.
to the request
The product has been added to the request.
to the request

Do you have specific requirements?
Select the appropriate attributes in our configurator and receive a customized quote for your request.

Configure product

Details

Why it is worth measuring compressed air flow rate

Consumption measurement shows the quantities required by individual machines, departments or production lines, and how compressed air consumption is distributed throughout the day. This allows costs to be allocated internally according to consumption and used as a key performance indicator in energy management in accordance with ISO 50001 or in energy audits in accordance with EN 16247-1. The base load at the end of a shift indicates leaks and continuous consumers; measurements taken before and after a measure is implemented show how effective it has been. Load profiles showing peak and base loads also provide a data foundation for planning pipework, dryers and compressors. The VC 600 enables this consumption measurement even on existing pipework without any intervention in the pipework, as it measures from the outside.

How clamp-on ultrasonic measurement works for compressed air and gases

The VC 600 is an ultrasonic flow meter based on the time-of-flight principle. ‘Clamp-on’ means that the sensor measures from the outside – without any intervention, often referred to as non-invasive. Two piezo transducers in the sensor unit alternately transmit ultrasonic signals through the pipe wall, both with and against the flow. The flow velocity is calculated from the difference in transit time; the volume flow rate is then derived from the velocity and the pipe cross-section. The difficulty with gases is that they conduct sound significantly less well than water; consequently, only a small proportion of the acoustic energy passes through the pipe wall into the gas. The VC 600’s sensor unit is therefore calibrated to the pipe wall thickness, ensuring that a sufficient signal is received even at low pressures; from 3 bar(g) upwards, the device measures reliably in metal pipes. Interference noise travelling through the pipe wall is absorbed by the damping mat in the quick clamp system.

Installation using half-shells and quick-release clamps without interfering with the pipework

The quick clamp system, made from 1.4404 stainless steel, consists of two half-shells with an integrated damping mat. It is placed around the pipe at a paint-free section and secured alternately using the four quick-release clamps; the clamps pull the half-shells evenly against the pipe so that the transducers make full contact. The sensor unit is then fitted and secured with two M4 screws. Ultrasonic gel or coupling pads are not required; the distance between the transducers is determined by the sensor unit, and no more than a screwdriver is needed.

Select the sensor unit and quick clamp system for pipes from 25 to 90 mm

The quick clamp system is selected according to the outside diameter of the pipe, in ten sizes ranging from 25 to 90 mm (¾″ to 3″). The sensor unit is available in two versions, designed for wall thicknesses of 1.5 to 2.2 mm and 2.2 to 5.0 mm; both fit all sizes of the quick-clamp system. Anyone wishing to measure several pipe sizes therefore only needs additional quick-clamp systems, not a second device. The outside diameter is the decisive factor, not the nominal diameter: A pipe with a nominal diameter of 50 mm has a diameter of 60.3 mm, a ‘50’ aluminium pipe has a diameter of 50 mm, and a stainless steel pipe in accordance with DIN EN 10357 has a diameter of 53 mm. Consequently, only the quick clamp system includes the pipe size code in the order reference.

Commissioning: Calculate standarized volume flow directly in the device

A set-up wizard guides the user through entering the inner diameter, wall thickness, pressure and temperature; pressure and temperature can be fixed values or continuously supplied by external sensors. Finally, the wizard checks the signal quality and confirms it. The VC 600 automatically converts the data to the selected reference conditions – ex-works in accordance with ISO 1217 – and, as a compressed air meter, totals the consumption; the CS SensorApp displays live values and trends via Bluetooth.

Retrofitting flow measurement during operation

Typical points of measurement for retrofitting include the supply line behind the dryer, branch lines to halls or cost centres, the supply line to a machine with conspicuous consumption, and high-pressure lines behind a booster; nitrogen consumption can also be allocated to individual consumption points in this way. The pipe remains closed; no new connection coming into contact with the medium is created. The measured values are transmitted via the selected interface to a chart recorder such as the DS 500 or directly to the energy management system.

Frequently asked questions about clamp-on flow measurement with the VC 600

FeatureVC 600
Measuring principleultrasonic transit-time difference, clamp-on
Measuring range0.08–20 m/s (F1); 0.08–25 m/s (F2)
Accuracy±1.5 % of reading ±0.3 % of full scale
Pipe outer diameter25–90 mm (¾″–3″); larger nominal sizes on request
Wall thickness1.5–5.0 mm (measuring unit B1 1.5–2.2 mm, B2 2.2–5.0 mm)
Pipe materialssteel, stainless steel, aluminium
Line pressurefrom 3 bar(g); upper limit set by the pipe, not by the sensor; suitability for high-pressure lines assessed from pipe and operating data
Gasescompressed air, nitrogen; others on request
Calibrationfactory calibration with certificate; standard via gas constant (D1), true gas calibration for the selected gas (D2); recalibration by sending in the measuring unit
Temperatureline −30 … +80 °C, ambient −20 … +70 °C
Reference conditionsISO 1217 (20 °C, 1,000 mbar, dry), DIN 1343 normal conditions for Nm³ (0 °C, 1,013.25 mbar), 15 °C/981 mbar, 15 °C/1,013.25 mbar (gas-industry base, ISO 13443)
Inputspressure 4–20 mA or Modbus RTU (PTS 500); temperature Pt100/Pt1000 or 4–20 mA
Outputs per versionI1: 4–20 mA + RS 485 Modbus RTU · I2: 4–20 mA + IO-Link · I3: Ethernet with PoE, Modbus TCP + 4–20 mA + RS 485 · I4: M-Bus + 4–20 mA + RS 485 Modbus RTU; Bluetooth for the CS SensorApp
Display, memorytouchscreen; totaliser; data logger 50 MB (about 2 years at 10 min)
Supply, protection, location18–36 VDC, 7 W; IP 65; indoor use, not for explosive atmospheres (Ex areas)

 

Examples for compressed air at 7 bar(g) and 20 °C, standard volume to ISO 1217, full scale at 20 m/s. The complete table for every clamping system is in the data sheet.

Inner diameterMeasuring range at 7 bar(g)
21.7 mm0.85–213 m³/h
41.9 mm3.18–794 m³/h
50.0 mm4.52–1,130 m³/h
60.3 mm6.58–1,644 m³/h
80.9 mm11.84–2,960 m³/h

A thermal sensor protrudes into the pipe and therefore requires a ball valve or a measuring section; the VC 600 is mounted on the outside of the pipe. It measures the flow velocity across the pipe cross-section and converts this to standard volume using pressure and temperature; a thermal sensor measures the mass flow directly and must therefore know the type of gas. 

Most clamp-on flow meters are designed for liquids. Gases conduct sound much less effectively, so a sensor unit is required that can transmit a strong enough signal through the pipe wall. The VC 600 has been developed for this purpose and measures flow in metal pipes from 3 bar(g) upwards.

The minimum pressure is 3 bar(g). The upper limit is not determined by the sensor but by the pipework, as the device is mounted externally and is not subjected to pressure. As the measuring range is specified in terms of actual flow velocity, the VC 600 measures high standardized volume flows at high pressures: At 40 bar(g) and an inner diameter of 50 mm, 20 m/s corresponds to approximately 5,800 m³/h, based on 20 °C and 1 bar absolute, using the ideal gas approximation. 

All three, each with its own function. The outside diameter determines the quick clamp system; you measure it using a calliper or by calculating the circumference (circumference divided by 3.14). The wall thickness determines the version of the measuring unit – 1.5 to 2.2 mm or 2.2 to 5.0 mm – and is included in the calculation; if it is unknown, the CS 0495 thickness meter measures it from the outside of the fitted pipe. The inner diameter is calculated as the outside diameter minus twice the wall thickness. It determines the pipe cross-section, from which the VC 600 performs the calculation for volume flow and consumption, and must therefore be correct: In the case of a DN 50 steel pipe (60.3 × 3.6 mm), underestimating the wall thickness by just 1 mm distorts the volume flow by around 8 per cent. The nominal diameter alone is not sufficient; the pipe material and pipe standard help to determine the outside diameter and wall thickness with certainty.

In compressed air networks, a relative pressure sensor such as the PTS 500 or CS 16 is sufficient: The VC 600 converts a fixed air pressure value of 1,000 or 1,013.25 mbar to the absolute pressure required for the standard volume calculation. A change in air pressure of 30 mbar alters the standard volume at 7 bar(g) by only about 0.4 per cent. An absolute pressure sensor can also be connected and adjusted on the device.

This is because the pipe wall determines how much sound reaches the gas. The sensor unit is calibrated for a specific wall thickness range to ensure that a sufficient signal is received even at low pressure. One version covers 1.5 to 2.2 mm, the other 2.2 to 5.0 mm. Both fit all quick-clamp systems.

Part of the ultrasound signal travels further into the pipe wall and is reflected back at flanges, welds and bolted joints. The damping mat in the quick clamp system absorbs this structure-borne noise at the sensor unit. If a flange is situated close to the point of measurement or if the pipe is very rough, this can affect the signal. In such cases, damping clamps fitted upstream and downstream of the sensor – suited to the pipe size and no more than 50 mm from the quick clamp system – can help.

For steel, stainless steel and aluminium pipes with an outside diameter of 25 to 90 mm and a wall thickness of 1.5 to 5.0 mm. Aluminium pipe systems for compressed air with outside diameters of 25 to 80 mm fall within this range. The installation site should be free of paint, rust and weld seams.

±1.5 per cent of the measured value ±0.3 per cent of the upper range value, i.e. the same value as for the VA 550 and VA 570 thermal sensors in their standard versions. This is subject to the pipe data being entered correctly and the point of measurement complying with the installation conditions.

It displays this as the base load: any current remaining at the end of a shift is attributable to leaks and continuous consumers. Where residual quantities are very small, we will agree the measuring range and measurement uncertainty with you. The individual leak can then be located using the LD 500 leak detector or the LeakCam 600 acoustic camera.

Every VC 600 is calibrated on the CS INSTRUMENTS volume flow test bench prior to dispatch and is supplied with a calibration certificate. For recalibration, only the sensor unit needs to be sent in; the quick clamp system remains attached to the pipe.

Product configurator

You can easily create one or more configurations for this product. 

When you add the product to the inquiry, all of your configurations will be automatically included in the inquiry.

Your current configuration

[[ error ]]

Select a variant

[[ completedSteps ]] / [[ totalSteps ]]
  1. [[ step.txt ]]

You have successfully completed your configuration.
Add it to your request now.

You have successfully completed your configuration.
You can now easily request it directly.

to the request

Not sure which option is right for you? Our experts will be happy to help.
Add the product, including all configurations, directly to your request. 

Not sure which option is right for you? Our experts will be happy to help.
Request the product — including all configurations — now.

to the request
[[ 'Your product is %xx% % configured'.replace('%xx%', progressPercentage.toFixed(0)) ]]

Related products

VA 500 - Flow meter for compressed air and gases

RS 485 interface, Modbus RTU, 4...20 mA analogue output for m³/h, Pulse output for m³.

More

VA 550 - Thermal mass flow sensor for flow measurement

Insertion sensor for demanding industrial applications, comes with ATEX + IECEx approval for Ex zones or DVGW approval for use in natural gas.

More

VA 520 - Thermal mass flow meter for flow measurement

for compressed air and gases. New digital electronics - improved accuracy.

More

VA 570 - Thermal mass flow sensor for flow measurement

ATEX + IECEx approval for Ex zones and DVGW approval for use in natural gas.

More

VA 526 – Compact flow sensor for compressed air and gases

Combines flow measurement, consumption monitoring, temperature measurement and, optionally, pressure and humidity/dew point monitoring in a single sensor.

More

VD 500 - Flow sensor for wet compressed air

For measuring immediately downstream of the compressor in moist air up to +180 °C

More

Accessories

Intelligent chart recorder DS 500 for compressed air and gases

Up to 12 optional sensors can be connected. 7" colour screen with touch panel.

More

PTS 500 - for measuring of pressure and temperature

2 in 1 sensor, measures pressure and temperature in the process.

More

Standard pressure sensor CS 10 and CS 16

for measurements from 0...10 and 0...16 bar with accuracy of full scale max. 1% or max. 0,5%.

More

Digital pressure sensor DPS 16

for measurements from 0...16 bar, RS485, G 1/2".

More

Thickness meter CS 0495

enables a quick, easy and accurate measurement of the wall thickness of pipes.

More

LeakCam 600 – all relevant leaks at a glance

Detects multiple leakages simultaneously with globally unique Power Beam Forming. 4th generation technology - also suitable for other gases. Helps to locate and repair leaks - including cost estimation and ISO 50001 documentation.

More