Details
Why measuring compressed air flow pays off
Compressed air submetering shows how much air individual machines, departments or production lines use and how compressed air consumption is distributed over the day. Costs can then be allocated internally by consumption and used as a KPI in ISO 50001 energy management or in a plant energy assessment. The baseline after the end of a shift points to leaks and permanent consumers; measurements before and after a measure show what it achieved. Load profiles with peak and base load are also the data basis for sizing piping, dryers and compressors. Because it measures from outside, the VC 600 makes this consumption measurement possible on existing piping without any intervention in the pipe.
How clamp-on ultrasonic measurement works on compressed air and gas
The VC 600 is an ultrasonic flow meter working on the transit-time (time-of-flight) principle. Clamp-on means the sensor measures from outside the pipe – non-invasively, also called non-intrusive. Two piezo transducers in the measuring unit send ultrasonic signals through the pipe wall, alternately with and against the flow. The transit-time difference gives the flow velocity; velocity and pipe cross-section give the volume flow. The difficulty with gases: they conduct sound far worse than water, so only a small part of the acoustic energy passes through the pipe wall into the gas. The measuring unit of the VC 600 is therefore tuned to the wall thickness of the pipe so that enough signal arrives even at low pressure; from 44 psig (3 bar(g)) the instrument measures reliably in metal pipe. Structure-borne noise traveling in the pipe wall is absorbed by the damping mat in the quick-release clamping system.
Mounting with half-shells and quick-release fasteners, without touching the pipe
The quick-release clamping system in 316L stainless steel (1.4404) consists of two half-shells with an integrated damping mat. It is placed around the pipe at a paint-free spot and closed with the four quick-release fasteners, alternating sides; the fasteners pull the half-shells evenly onto the pipe so that the transducers couple in with full contact. The measuring unit is then set on top and secured with two M4 screws. No ultrasonic gel or coupling pads are required, the spacing between the transducers is fixed by the measuring unit, and nothing more than a screwdriver is needed. Mounting location and straight pipe runs are part of choosing the measuring point; the details are in the operating instructions.
Selecting the Measuring Unit and Clamping System for Pipe Outside Diameters of 0.98 to 3.54 in. (25 to 90 mm)
Three pipe dimensions work together. The outside diameter determines the quick-release clamping system: ten sizes cover 0.98 to 3.54 in. (25 to 90 mm) OD, so every steel pipe from NPS ¾ to NPS 3 has a matching clamping system. The wall thickness determines the measuring unit version: one for 0.059 to 0.087 in. (1.5 to 2.2 mm), one for 0.087 to 0.197 in. (2.2 to 5.0 mm); both fit every clamping system. For steel pipe, Schedule 40 is within range up to NPS 2, Schedule 10 up to NPS 3. Outside diameter minus twice the wall thickness gives the inside diameter, which you enter on the device or in the app – the VC 600 calculates the flow from it. Anyone who wants to measure several pipe sizes therefore needs only additional clamping systems, not a second instrument, provided the wall thicknesses fall in the same range. What counts is the outside diameter, not the nominal size: a 2-in. NPS steel pipe measures 2.375 in. (60.3 mm) outside, a 2-in. sanitary stainless tube 2.0 in. (50.8 mm), and aluminum piping systems are often specified by metric OD such as 50 or 63 mm. In the order code, the clamping system carries the pipe-size code and the measuring unit the wall-thickness code.
Commissioning: calculating standard flow in the device
A setup wizard guides the user through entering inner diameter, wall thickness, pressure and temperature; pressure and temperature can be fixed values or come continuously from external sensors. Finally the wizard checks the signal quality and confirms it. The VC 600 converts to the selected reference conditions itself – ISO 1217 from the factory, i.e. 68 °F and 14.5 psia (20 °C, 1,000 mbar) – and totalizes consumption as a compressed air meter; via Bluetooth the CS SensorApp shows live values and trends.
Retrofitting flow measurement while the plant is running
Typical measuring points for a retrofit without shutdown are the supply line after the dryer, branches to individual buildings or cost centers, the feed line to a machine with conspicuous consumption and high-pressure lines after a booster; nitrogen consumption can be assigned to individual points of use in the same way. The line stays closed, and no new connection in contact with the medium is created. The readings go via the selected interface to a data logger such as the DS 500 or directly to the energy management system.